Comparison of Euroimmunn line blot and immunoprecipitation testing for anti-Mi2 autoantibodies. This study found that only subjects that are positive for both anti-Mi2alpha and and-Mi2beta autoantibodies by line blot are likely to positive by IP. Those that are positive for only anti-Mi2beta are generally false positives and 50% of the anti-Mi2alpha only are false positives.