Performance evaluation of three assays for the detection of PR3-ANCA in granulomatosis with polyangiitis in daily practice

Rate this item
(0 votes)

Utility:

Consecutive patient sera that tested ANCA IFA positive was tested with an EUROIMMUN ELISA, utilizes native and recombinant PR3 protein, and Inova QUANTA LITE and QUANTA Flash assays that utilize only native PR3 antigen. Nineteen of the 78 patients had GPA and were identified by all three assays; however, the specificity of the EUROIMMUN ELISA was the lowest (61%). The QUANTA lite had the highest specificity (81.4%) and likelihood ratios (5.36) and the QUANTA Flash assay had a specificity of 69.5% and likelihood ratio of 5.36. The Burmingham Vasculitis Score (BVAS) is a measure of disease severity. Both the QUANTA Lite and QUANTA Flash PR3 titers correlated with BVAS, whereas, the EUROIMMUN assays did not and is likely a result of the low-binding specificity of non-pathogenic antibodies.