Performance evaluation of a novel chemiluminescence assay for detection of anti-GBM antibodies: an international multicenter study

Rate this item
(0 votes)

Utility:

The performance characteristics of the QUANTA Flash anti-glomerular basement membrane (GBM) assay was assessed in patients with Goodpasture's syndrome (GPS). The study showed the sensitivity and specificity of 95.6% and 99.6%, respectively. When the results were compared to other methods, good agreement was observed: 95.8% vs. EliA GBM, 97.4% vs. both BINDAZYME Anti-GBM and QUANTA Lite GBM.